Intended use & safety
Supplied by Synocular Ltd · Product: Synocular eye tracker (illuminator bar) · USB-C and 5 V DC input. Please read this notice before using the product. It sets out the research-and-development, business-to-business basis on which Synocular is supplied.
What this product is
The Synocular eye tracker is being developed for professional research use. It is not currently available for supply. When available, it will be supplied business-to-business for research and development use. It is not intended for diagnosis, screening or treatment, and is not supplied for consumer, home or clinical use.
Infrared / eye safety
This device contains six 850 nm near-infrared (IR) LED emitters, in two groups of three. Their exposure has been calculated against the IEC 62471 Exempt-group limits: the calculated corneal irradiance at 200 mm is about 44 W/m², against a limit of 100 W/m². An independent photobiological safety assessment against IEC 62471 is planned before supply; final safety information will be published when it is complete.
- Do not place your eye closer than 200 mm from the emitters. This is the design minimum, which the independent assessment will confirm. The product is intended for use with the eyes 55–75 cm from the tracker, which sits just below the screen.
- Do not view the emitters at close range through magnifying optics.
- The light is largely invisible and there is no blink reflex to warn you, so observe the minimum distance above.
- Not intended for use on, or unsupervised near, infants or young children.
Electrical
Powered by 5 V DC only (no mains voltage enters the product): USB-C for data, and a separate isolated 5 V DC supply (4.75–5.20 V, rated 3 A or more, 5.5 × 2.1 mm plug, centre positive) for the illumination, which draws up to 1.8 A. Use the supplied power supply and cabling. A replacement supply must be isolated and meet the same rating.
Electromagnetic compatibility (EMC)
Each tracker will be supplied with a declaration of conformity under the UK Electromagnetic Compatibility Regulations 2016, once EMC testing is complete (planned for early 2027).
Mount (if supplied)
The mount is 3D-printed and fixed with screws, nuts and hand knobs. Keep small parts away from young children and anyone who might swallow them.
Not a medical device
This product is not a medical device and must not be used for diagnosis, screening, treatment, or any clinical purpose.
Where the Synocular gaze engine or tracker is built into a communication aid, computer-access product or other finished product, that product is placed on the market by its maker, who is responsible for its intended purpose, its conformity assessment and any medical-device registration it needs. Synocular supplies the component and its technical information.
Disposal (WEEE)
Do not dispose of in general household waste. Dispose of as electrical/electronic equipment in accordance with local regulations.
Conditions of supply (summary)
- Supplied business-to-business for research and development use, with a 12-month return-to-base warranty against faults in materials and workmanship, and without warranty of fitness for any particular purpose beyond research and development use.
- The user is responsible for setting up and using the tracker in line with this notice.
- These conditions do not exclude any liability that cannot lawfully be excluded.
Synocular Ltd, 71-75 Shelton Street, Covent Garden, London WC2H 9JQ, United Kingdom · synocular.com · contact@synocular.com. This notice is supplied with every unit.